Voyager - Mobile Particle Size Analyzer

On-line and at-line mobile particle size analyzer:

  Solid or liquid particulates suspended in a gas process stream
 
  Automatic on-line continuous, or at-line manual
 
  Mobile installation
 
  Real-time particle size analyzer based on laser diffraction 0.1 to 2500µm
 
  Ideal for performing consultancies, trouble shooting or process optimization
 

Award winning technology

  • Most Innovative New Product
    Asia Pacific Coatings Show 2003
  • Processing Magazine Breakthrough Product of 2003

Insitec technology for dry powder applications is now available as a neatly packaged mobile particle size analyzer that can be moved easily around the plant for real-time on-line particle size analysis, or for automated at-line measurement. Voyager is especially useful for sites with multiple unit operations such as mills, spray driers and classifiers.

The Voyager is a neatly packaged mobile at-line and on-line particle size analyser
The Voyager is a neatly packaged mobile at-line and on-line particle size analyser
Laser Diffraction
The particle size analyzer uses industry standard laser diffraction.


Process Interface
The Insitec uses a modular system for connection to many different types of process streams.


Automation & control
The automation of the analyzer system and connection to the plant control system is facilitated via Malvern Link, a flexible approach accepting custom user interfaces, and a variety of industrial communication protocols


Solutions
The particle size analyzer system has been proven in a variety of different industrial applications, from cement to pharmaceuticals.

 

What's Next?



 

Summary

The configuration can be specified to suit particular applications, however the base system is configured as follows:
 
 

General

Analyzer

Insitec T

  • IP66 rated
  • Dynamic Range 0.1 to 1000 µm
  • Power supplied from the PC cabinet
  • Integral window purge
  • Fully tri-clamped stainless steel optical bench, the ultimate in easy cleaning
Installation
Mobile unit
  • Permits easy transportation of entire measurement system.
  • Four puncture proof pneumatic wheels
  • Provides mount for Insitec Analyzer, Air Panel, Electronics Enclosure and Lap Top PC
  • Allows rotation of Insitec Analyzer to match process line orientation.
  • Insitec analyzer easily dismounted to facilitate connection to process line if required.
  • Rugged construction suitable for any Plant environment.
Process Interface Kit Standard on every system:
  • Dispersion provided by 90° Stainless Steel Venturi Eductor
    (alternate dispersion systems available as options)
  • 80cm Stainless Steel test flute with pre-drilled holes to allow rapid optimization of probe configuration.
  • 2 meters flexible inlet pipe
  • 2 meters flexible outlet pipe
  • 3 fixed sampling point flanges + 80 cm standard flutes
Standard system only:
  • Inlet and outlet ball valves with triclamp fittings (Manual)
Air supply

Air Panel Assembly

  • Automated air panel housed in IP65 polyester cabinet.
  • Low-flow purge to protect windows when system is idle.
  • Enables manual control of eductor and window purge flows. Housed in a lockable IP65 polyester cabinet with clear window.
  • Three adjustable air flow meters, automatic valves and flow sensors.
  • Boost air facility
  • Pinch valve control
  • All automatic functions configurable via Malvern Link
Electronic enclosure
  • Lockable IP65 rated cabinet - housing:
  • Embed PC to allow stand alone operation
  • Ethernet connection for configuration PC or plant control
  • Diagnostic LED's at the Optical sensor
  • Voltage is auto sensed in range 100 to 240 volts
Cables
  • 5m of cable to connect between Optical Head and Interface box.
  • 5m mains lead.

Automation

Automation cabinet

Voyager:

  • Hardware supporting full Malvern Link functionality housed in IP65, containing:
  • Wago, analog and digital output modules (3 A/O)
  • Power Supply
  • Switches for installation testing
  • Break outbox
  • Control for external vacuum system
  • Click here for more details
Software Suite

The following software is pre-installed as standard with each process system (and also on CD-ROM). They are compatible with Microsoft Windows NT4.0 and XP for use on Intel compatible PC and available in English, French, German, Spanish and Japanese languages.

  • RTSizer for instrument configuration, data acquisition and particle size analysis; includes high concentration software module, patented technique to enable measurement under high or fluctuating process loadings (5-95% transmission)

  • Malvern Link for instrument control and operator interface, including license for award winning Intellution iFix SCADA software.


Malvern Link Software and Computer

Voyager:

  • Malvern Link options such as Plant Optimizer, Batch Tracker...Access to Engineering Mode of Malvern Link.

Option:

  • "Toughnote" IP54 laptop computer


Miscellaneous

Installation & Training

  • Includes two days training at Malvern office plus three days on-site training. Upon completion of training, Malvern Process Systems will offer on-site applications assistance at a 10% discount to our standard daily rate for the first twelve months. T&E charged at cost.

  • Installation, commissioning and additional training days excluded. User is responsible for installation, commissioning and application development.

Site requirements

  • Location for instrument to be specified (following consultation from Process Specialist engineers):

  • On-line (eductor) configuration: Sample ports (supplied by Malvern Process Systems) to be connected to process line

  • In-line (flowcell) configuration: Process line broken and flanges added to connect to instrument.

  • A dry clean air or nitrogen supply of a minimum of 30m3/hour at 4 bar g (30scfm at 60 psi g) is required at instrument location to provide for window purge and motive air.

  • 100-240 Volts AC, 16 AMPS and ground connection is required on site to provide power for sensor and interface enclosure.

 


Specification

Lens size range

Choice of four size ranges (one lens is delivered as standard, others can be ordered as options):

Lens focal length (mm)

Size range (µm)

100

0.1 - 200

200

1.0 - 400

300

1.5 - 600

450

2.25 - 1000


Materials
  • 316L for all metallic parts, with high grade finish
  • Glass windows (for measurement zone)
  • Viton O-rings and EPDM triclamp gaskets (other seal materials can be supplied)


Fittings

Triclamped for simplified cleaning

Transmission

5 - 98 %

Accuracy

±2% on Dv(50) reported using the verification reticle

Maximum operating pressure

10 bar (g)

Operating temperature (ambient)

-10°C to +50°C

Industrial protection

Dust-tight and waterproof: Optical head IP66, cabinets IP65

Regulatory compliance

Validation and 21 CFR part 11 options (If required)

Automation

Voyager contains:

  • valve sequencing for automatic background, sample path cleaning, start / stop
  • OPC server to plant DCS
Dimensions

Height range 0.5m to 1.5m x W 1.2m to 1.6m x D 0.97m

Weight (approx)

Insitec 25kg

Trolley 30kg

Automation 20kg