Mie scattering (Laser diffraction).
|
|
|
| |
Laser Diode 670 nm wavelength
|
|
|
| |
10 mm
6 mm
3 mm 1/e2
|
|
|
| |
1 to 4 Hz.
|
|
|
| |
Sprays with obscurations from 2% to 95%.
|
|
|
| |
3% (using RS-3 reference standard reticle).
|
|
|
| |
0.5% (using RS-3 reference standard reticle).
|
|
|
| |
Optics can be purged with clean air to protect them from contamination during measurement. A minimum supply of 2ft³ min. is required - this air must be completely oil and water free.
|
|
|
| |
Basic, IP65(Protected against dust / water jets. European standard).
|
|
|
| |
Pentium II 300 MHz processor, 32 Mb RAM, 3Gbyte hard disc and CD–ROM.
Screen resolution VGA 800x600 (SVGA 1024x768 is recommended)
Requires 1 free serial port.
|
|
|
| |
Windows 2000
|
|
|
| |
Model-independent analysis gives droplet size distribution Patented Multiple Scattering Algorithm for operation at high concentrations (US Patent No. 5,619,324).
Statistical analysis for QA/QC.
|
|
|
| |
User configurable SPC charting and report formats.
|
|
|
| |
"Frame-by-frame" data display.
|
|
|
| |
Direct to Excel or similar spreadsheet.
|
|
|
| |
Transmitter module 5.75 kg.
Receiver module 7.50 kg.
|
|
|
| |
Short Bench - 300 mm spray gap. Height of laser above bench 250 mm
Long Bench - 500 mm spray gap. Height of laser above bench 250 mm
Custom Benches are available at extra cost depending on material, finish and size.
|
|
|