Summary
The Parsum probe is inserted directly into a process line with the following general configuration and parameters.

|
|
General |
|
Intrinsic safety |
- The IPP 70-SE unit IS Intrinsically Safe
- Explosion Group Class IIB
- Temperature Class T4, with a temperature range of -20°C to 40°C
- Casing Protection IP65
|
|
Sensor
|
Applies an extended fibre-optic local filter method |
|
Installation
|
Is directly inserted into process line in the hazardous area with a cable to the measurement computer in the explosion free zone . |
|
Process interface kit
|
- Weld Stub to be installed on the process line
- Feed-thru piece to hold probe
- Tri-clamp to hold probe to the feed-thru piece
|
|
Air supply
|
The IPP 70-SE unit requires dust free instrument grade compressed air with a maximum flowrate of 25L/min. |
|
Cables
|
- 100m maximum for the IS cable. 100m from the feeding devices to the PC
|
|
Site Requirements
|
- Installation location to be specified following consultation of a Malvern Process Specialist
- Welding the support stub in the process line
- Conduit laid for cables to computer and to control room
- Minimum 120 V AC 5amp power supply
|
|
| Software and Hardware |
- Inline Particle Probe Software V. 7.1
- NI 5102 A/D Converter card or NI USB-5133 optional
- Field Mounted Computer with Windows 2000 Operating System
|
|
| Feedback Control Options |
- PCI -6308A D/A Converter Card (4-20 mA Outputs)
- RS485 Interface Card
|
|
Miscellaneous |
|
Installation & Training |
|
|
Standard documentation |
The following documents are supplied on a CD-ROM in PDF format and are available in English, German, French and Spanish languages:
- Commissioning Manual
- Operator Control Software Manual
|
|
Specification |
|
| Measurement range |
- Particle Size Measuring Range 50-6000µm.
|
|
| Intrinsically safe |
The IPP 70-SE unit is intrinsically safe.
|
|
| Materials |
- 316L SS for in-line probe.
- Sapphire, epoxy resin optics.
- Pressure-cast aluminum for casing.
|
|
| Type of measurement material |
Solid, liquid and gaseous particles.
|
|
| Concentration |
200-120,000 ppm volume content, dependent on particle size and velocity of the moving particles.
|
|
| Air Requirements |
The IPP 70SE unit requires dust free instrument grade compressed air with a maximum flowrate of 20 L/min.
|
|
| Maximum Operating Pressure |
4 bar at the probe's sensing volume.
|
|
| Operating Temperatures |
- -20°C to 85°C at the measuring point
- -10°C to 60°C on the probe housing
|
|
| Industrial Protection |
IP 65 (probe)
|
|
| Maximum Number of Instruments Connected to one PC |
4
|
|
| Dimensions |
- Tube Length = 280mm (~11in).
- Tube Diameter = 25mm (~ 1in).
|
|
| Weight (approx) |
1.5kg
|
|
|
| |