Summary
The configuration can be specified to suit particular applications, however the base system is configured as follows:
|
|
|
General |
|
|
Intrinsic safety |
Non-IS option: Insitec T
IS option: Insitec X
|
|
|
Analyzer |
Non-IS: Insitec T
- IP66 rated
- Dynamic Range 0.1 to 200 µm
- Power supplied from the Human Machine Interface (HMI) PC cabinet
- Integral window purge
- Fully tri-clamped stainless steel optical bench, the ultimate in easy cleaning
- Highly polished monoblock flow cell
IS: Insitec X
- IP66 rated
- Dynamic Range 0.1 to 200 µm
- Power supplied from the Human Machine Interface (HMI) PC cabinet
- Integral window purge
- Fully tri-clamped stainless steel optical bench, the ultimate in easy cleaning.
- Highly polished monoblock flow cell.
|
|
|
Installation
|
Mobile unit
- Permits easy transportation of entire measurement system.
- Purged enclosure to meet area classifications
- Four wheels, two rotatable (for steering)
- Provides mount for Insitec Analyzer, Air Panel, Electronics Enclosure and Lap Top PC
- Allows rotation of Insitec Analyzer to match process line orientation.
- Insitec analyzer easily dismounted to facilitate connection to process line if required.
- Removable transport panels allow easy access to Insitec Analyzer and controls.
- Rugged construction suitable for any Plant environment.
- Easily shipped to any Plant in the world.
|
|
|
Air supply
|
Air Panel Assembly
- Three adjustable air flow meters (lockable to prevent inadvertent adjustment)
- Low-flow purge to protect windows when system is idle
- Cartridge filters
|
|
|
Electronic enclosure
|
- Stainless Steel nema 4x Enclosure with Pedestal and elastomer keyboard
- Communication interface PCB - allows RS232 or RS422 connection to the PC.
- Maximum cable length is 15m using RS232. The RS422 connection should be used for longer length requirements.
- Diagnostic LED's at the Optical sensor
- Voltage is auto sensed in range 100 to 240 volts
|
|
|
Cables
|
- 10m of cable to connect between Optical Head and Interface box.
|
|
|
Software |
|
|
| Software Suite |
The following software is pre-installed as standard with each process system (and also on CD-ROM). They are compatible with Microsoft Windows NT4.0 and 2000 for use on Intel compatible PC and available in English, French, German, Spanish and Japanese languages.
-
RTSizer for instrument configuration, data acquisition and particle size analysis; includes high concentration software module, patented technique to enable measurement under high or fluctuating process loadings (5-95% transmission)
-
Malvern Link for instrument control and operator interface, including license for award winning Intellution iFix SCADA software.
|
|
|
| Malvern Link Software and Computer |
Industrial PC supplied inbuilt to voyager with elastomer keyboard .
|
|
|
Miscellaneous |
|
|
Installation & Training |
-
Includes two days training at Malvern office plus three days on-site training. Upon completion of training, MALPS will offer on-site applications assistance at a 10% discount to our standard daily rate for the first twelve months. T&E charged at cost.
-
Installation, commissioning and additional training days excluded. User is responsible for installation, commissioning and application development.
|
|
|
Standard documentation |
The following documents are supplied on CD-ROM in PDF format and available in English, French, German, Spanish and Japanese languages.
- Material certificates
- Installation manual
- User manuals
|
|
|
Site requirements |
-
Location for instrument to be specified (following consultation from MALPS engineers):
-
On-line (eductor) configuration: Sample ports (supplied by MALPS) to be connected to process line
-
In-line (flowcell) configuration: Process line broken and flanges added to connect to instrument.
-
A dry clean air or nitrogen supply of a minimum of 50m3/hour at 4 bar g (30scfm at 60 psi g) is required at instrument location to provide for window purge and motive air.
-
A fused spur rated at 100-240 Volts AC, 5 AMalPS and ground connection is required on site to provide power for sensor and interface enclosure.
-
Local electricity supply
-
Instrument - PC data communications (cable supplied with instrument) to be installed from instrument location to PC location
|
|
|
Specification |
|
|
| Lens size range |
Choice of four size ranges (one lens is delivered as standard, others can be ordered as options):
Lens focal length (mm) |
Size range (µm) |
100 |
0.1 - 200 |
200 |
1.0 - 400 |
300 |
1.5 - 600 |
450 |
2.25 - 1000 |
|
|
|
| Intrinsically safe |
Depends on optical head chosen.
Insitec T : NOT intrinsically safe
Insitec X : EEx ia IIC T4.
Approved for use in zones 0, 1 and 2 (gas environments) and zones 20, 21, 22 (dust environments)
|
|
|
| Materials |
- 316L for all metallic parts, with high grade finish
- Glass windows (for measurement zone)
- Viton O-rings and EPDM triclamp gaskets (other seal materials can be supplied)
|
|
|
| Fittings |
Triclamped for simplified cleaning
|
|
|
| Transmission |
5 - 98 %
|
|
|
| Accuracy |
±2% on Dv(50) reported using the verification reticle
|
|
|
| Air requirements |
Dry, oil-free air containing no particulates
6 bar g (90 psi g) with a flowrate of 30 m³/h (1000 ft³/h) required as motive gas for eductor and for window purge
Note: other gases (e.g. nitrogen) can be used instead of air as appropriate
|
|
|
| Maximum operating pressure |
10 bar (g)
|
|
|
| Operating temperature (ambient) |
-10°C to +70°C
|
|
|
| Industrial protection |
Dust-tight and waterproof: Optical head IP66, cabinets NEMA 4x
|
|
|
| Maximum distance from instrument to PC |
500 m
|
|
|
| Maximum number of instruments connected to one PC |
1 |
|
|
| Software |
RTSizer (for instrument control)
Malvern Link (for system automation and data link)
All software runs on MS-Windows 2000/XP or above
|
|
|
| Regulatory compliance |
Validation and 21 CFR part 11options
|
|
|
| Dimensions |
H 1.3m x W 0.9m x D 0.7m
|
|
|
| Weight (approx) |
165kg
|
|
|
| |