Skip to left navigation

Skip to page content

Find us on... Facebook Twitter Materials Talks Blog Google Plus Linked In YouTube

Dual Integrated Light Scattering system dedicated to protein characterization. Use as a chromatography detector with any GPC/SEC system to give absolute size and absolute molecular weight. Use in cuvette mode for the detection of large aggregates.

Overview

Dynamic light scattering technology (DLS) used to measure molecule size, and Static Light Scattering (SLS) to determine molecular weight.

  • Use as a chromatography detector to measure absolute size and absolute molecular weight, aggregation and conjugation by connecting the Zetasizer µV to your GPC/SEC system and concentration detector, using standard chromatography tubing.
  • Use in cuvette mode - simply remove the flow cell from the optics unit and insert a cuvette containing the sample.
  • Switch in seconds from chromatography detector to cuvette mode.
  • Accurate and repeatable results using precise temperature control.
  • Sensitivity – 0.1mg/mL 15kDa protein (Lysozyme).
  • Only 2µL sample required in cuvette mode to preserve sample.
  • 8µL flow cell for chromatography mode ensures minimal band broadening to preserve the resolution of the separation.
  • Connect to any Viscotek system for full software integration and truly automated operation, or third party SEC system using OmniSEC software and OmniFACE interface.
  • Automated temperature trends for aggregation temperature determination (melting point).

Back

How it works

The Zetasizer µV incorporates two techniques that can be used in both chromatography detector mode and cuvette mode.

Dynamic Light Scattering is used to measure particle and molecule size. This technique measures the diffusion of protein molecules moving under Brownian motion, and converts this to size and a size distribution using the Stokes-Einstein relationship.

Static Light Scattering is used to determine molecular weight. When used as a detector, the light scattering intensity in combination with the signal from a concentration detector enables the absolute molecular weight to be calculated.

In cuvette mode, the intensity of a few concentrations are measured to enable a Debye plot to be constructed. From this a molecular weight and second virial coefficient can be calculated.

The software is designed to be versatile and rich in features without compromising ease of use. Standard Operating Procedures (SOP) simplify routine measurements, a quality factor gives a single place for confirmation of data quality, and an expert advice system is provided to assist with data interpretation.

Back

Specification

Measurement type 1
Particle size, molecular size
Measurement range:
0.3nm – 2.0 microns* (diameter)
Measurement principle:
Dynamic Light Scattering
Measurement angle
90°
Minimum sample volume:
2µL
Accuracy:
Better than +/-2% on NIST traceable latex standards
Precision/Repeatability
Better than +/-2% on NIST traceable latex standards
Measurement type 2
Chromatography detector, size, molecular weight
Measurement range
0.3nm – 2.0 microns* (diameter) column dependent
Measurement principle
Dynamic Light Scattering, Static Light Scattering
Measurement type 3
Molecular weight
Measurement range
< 1000Da – 20M Da.*
Measurement principle
Static Light Scattering using Debye plot
Minimum sample volume
2µL (3-5 sample concentrations required)
Accuracy
+/- 10% typical
Temperature control range
2°C - 90°C +/-0.1.**
Light source
Laser 830nm, Max 60mW
Laser safety
Class 1
Power (V, Hz, VA)
100VA
Dimensions (W, D, H)
350mm x 410mm x 170mm (W, D, H)
Weight
14kg
Operating environment
Temperature
15°C – 40°C
Humidity
35% - 80% non-condensing
Software
Recommended computer specification
Intel i5 processor PC, 4GByte RAM, 500GByte hard disk drive, 1280 x800 screen resolution (widescreen), or 1280 x 960 (4:3), CD-ROM drive, USB port.
Operating platforms
Windows XP (SP2), Windows 7 operating system recommended.
*
Maximum range, sample dependent.
**
At 25°C

Back

Accessories

Zetasizer µV including chromatography interface and software

System available including all hardware and software for connection to a third party SEC system.

Unit available alone for connection to a Viscotek SEC system.

21CFR part 11 software

Software option to enable compliance with electronic records and electronic signatures (ER/ES).

Totally integrated into the standard software to provide multiple level password controlled access to method changes, and a full audit trail.

Back

Malvern solutions for the Zetasizer µV



Introduction to the Zetasizer µV as a highly sensitive SEC detector The Zetasizer µV is a 90° light scattering instrument capable of both dynamic light scattering (DLS) and static light scattering (SLS) measurements. It has a...
Introduction to the Zetasizer µV as a highly sensitive SEC detector The Zetasizer µV is a light scattering system capable of simultaneous dynamic (DLS) and static light scattering (SLS) measurements of size and molecular weight. It can...
Measurements of 0.1mg/ml Lysozyme in 2µl cell volume Measurements made at the sensitivity limits of DLS instruments are challenging but good experimental practice and careful sample preparation coupled with appropriate...


 
 
Request a demo Request a quote Contact us
 

Free registration

Access application notes, webinars, videos and more.