Viscotek HT-GPC Light Scattering
High temperature GPC/SEC system, with light scattering detector and RI detector.
The Viscotek HT-GPC systems are designed specifically for the molecular weight and structure characterization of polyolefins, but with the ability to handle other polymers, such as polythiophene, PEEK and PVF that require temperatures in the region of 100°C to 160°C to stay in solution.
The Viscotek HT-GPC Light Scattering is capable of making the following measurement types:
- Relative Molecular Weight (Mn, Mw, Mw/Mn).
- Absolute Molecular Weight (Mn, Mw, Mz, Mw/Mn).
The OmniSEC software controls all the functions of the HT-GPC and provides the simplicity of two clicks from data to results. It also has the power to handle more complex calculations such as branching analysis (for systems fitted with a viscometer).
Overview
The HT-GPC Light Scattering system comes with a high temperature RI, for use as a concentration detector, and dual Light Scattering detectors (LALS and RALS) for absolute molecular weight determination.
It also features a temperature controlled column chamber and automated filtering of the sample at high temperature, to ensure safe handling.
- Dual piston optimized for HT-GPC, with extremely low pulsation and stable eluent flow to improve detector performance.
- Temperature controlled environment for up to 7 columns, injector and up to three detectors.
- Automated sample preparation using the Vortex autoprep/autosampler. Operates up to 160°C with no user intervention, making the system safe, accurate and productive.
- Automated sample filtration reduces the number of preparation steps – only the sample injected on the column is filtered, which avoids bulk filtration and improves lab safety (no manual handling of hot solvents).
- Unique 7° Low Angle (LALS) detector measures the absolute MW directly without data fitting, extrapolation or correction.
- 90° Right Angle (RALS) detector provides the best signal to noise ratio and the highest sensitivity for small molecules.
- Accurate measurement of the polymer concentration using a refractive index detector.
How it works
The HT-GPC pump passes mobile phase through the GPC/SEC chromatography columns, following by the injector. The columns are maintained at the same temperature as the injector and detectors. After eluting from the column, the dissolved polymer molecules, now separated by size, enter the light scattering and RI detectors. Light scattering detection is used to determine the absolute molecular weight of dissolve particles, and refractive index detection is used to measure sample concentration.
Specification
- Component:
- Oven/Detector module
- Temperature Control:
- 30°C to 160°C all detector cells and columns.
- Column Capacity:
- Up to 7 standard 30cm GPC/SEC columns.
- Interface:
- 6 channel digital interface to OmniSEC software.
- Inputs:
- 2 analog inputs for auxiliary detectors.
- Component:
- Vortex Autosampler
- Capacity:
- 30 vials, 40mL, (20mL option).
- Temperature Range:
- 30°C to 160°C.
- Stirring Speed:
- Variable, 0 – 300 rpm.
- Sample Filtration:
- Automatic self-cleaning in-line sample filtration.
- Sample Preparation:
- Optional semi-automated sample preparation.
- Component:
- Degasser
- Channels/volume:
- Two channels, 8mL per channel.
- Performance:
- <0.5ppm oxygen at 0.5mL/min.
- Component:
- Pump
- Flow rate range:
- 0.01 to 9.99mL/min.
- Pulsation:
- Less than 1% (measured with viscometer).
- Pressure Reading:
- MPa or PSI.
- Soft start/stop:
- User programmable in mL/min/min.
- Detector 1:
- Light scattering detector.
- Measurement principle:
- RALS, 90° angle; LALS, 7° angle, with high efficiency optics.
- Light source:
- Temperature controlled laser diode, 10mW, 660nm.
- Cell:
- 18µL.
- Purging:
- 3 modes, manual, automatic and programmable.
- Data rate:
- 100Hz collection, DSP with 5Hz to OmniSEC.
- Detector 2:
- Differential refractive index detector.
- Measurement principle:
- Deflection.
- Light source:
- LED, 660nm.
- Cell:
- 45° quartz glass, 12µL volume.
- Purging:
- 3 modes, manual, automatic and programmable.
- Data rate:
- 100Hz collection, DSP with 5Hz to OmniSEC.
- Measurement type 1:
- Absolute Molecular Weight
- Measurement principle:
- ight scattering (LALS or RALS).
- Measurement type 2:
- Concentration
- Measurement principle:
- Differential Refractive Index detection.
- Operating environment:
- Temperature:
- 15°C – 40°C.
- Humidity:
- 35% - 80% non-condensing.